A mean error of zero sounds excellent. A spectrum can still turn that respectable-looking average into a statistical practical joke.
Calculate signed mass error as observed minus reference m/z, divided by the reference m/z, multiplied by one million. Compare individual errors before interpreting the average. Positive and negative errors can cancel, so an average near zero can coexist with poor repeatability or unacceptable individual measurements.
Work out signed error for each reading
| Reference m/z | Observed m/z | Signed error |
|---|---|---|
| 2,000.000 | 2,000.040 | +20 ppm |
| 2,000.000 | 1,999.960 | −20 ppm |
| 2,000.000 | 2,000.000 | 0 ppm |
What would you report?
The mean signed error is 0 ppm. The maximum absolute error is 20 ppm, and the mean absolute error is about 13.3 ppm. If the exercise sets a ±10 ppm limit for each measurement, two readings fail even though the mean is zero. That limit is fictional; actual acceptance criteria come from the lab method and instrument configuration.
Run a thought experiment with +18, +19 and +20 ppm. The errors are tightly grouped but consistently positive, which points to an offset even though the measurements are more repeatable. Record the mode, expected masses and which independent checks were used before interpreting either pattern.
Read the MALDI-TOF calibration and troubleshooting guide · Example manufacturer specifications and conditions